Biochemical and Pain Comparisons Between the Laser Lancing Device and Needle... | Wiley Online Library
Writer LAMEDITECH Date 2024-03-15 Hit 317

 

Abstract

 


Background and Objectives

 

Patients around the world use a lancing device to perform self‐monitoring of blood sugar (SMBG). However, there are always fears of needles and pain. Therefore, less painful devices are being developed. The purpose of this study was to compare the usefulness and safety of a laser lancing device (without a needle) to a conventional needle lancet (with a needle) for capillary blood sampling.

 

 

Study Design/Materials and Methods

 

A total of 40 healthy subjects were enrolled in the study. Capillary blood was collected from a laser lancing device (without a needle) and a conventional needle lancet (with a needle) on opposite fingers, the choice of which was randomly selected. The laser lancing device (LMT‐3000) uses a 2940 nm mono‐pulse laser, a radiation field of 350 μm, laser energy of 210 mJ, and a 3.7 V battery. One week later, capillary blood was obtained by switching the devices and fingers. The biochemical measurements and pain were compared between the two groups. Puncture pain was measured on a pain scale from 0 to 10.


Result

 

All patients were tested with both a laser lancing device and a conventional needle lancet. In the biochemical analysis, the blood glucose level was 103.21 ± 17.20 mg/dl in laser lancing device group and 102.25 ± 22.44 mg/dl in the conventional needle lancet group, and there were no significant differences between the two groups ( = 0.940). The pH, CO2, O2, lactate and hematocrit levels of the blood were no significant differences between the two groups. In the first trial, the median pain score (interquartile range) of patients using laser lancing device was 2.0 (1.0–3.0), whereas it was 2.5 (2.0–4.0) in patients using a conventional needle lancet ( = 0.029). In the second trial, one week later, the median pain score in the laser lancing device group was 2.5 (1.0–4.0), whereas it was 3.5 (2.25–5.0) in the conventional needle lancet group ( = 0.001). The difference in pain scores between the first and second trials was significant in the conventional needle lancet group ( = 0.007), but not in the laser lancing device group ( = 0.150).            

 

 

Conclusion

 

There was no difference in biochemical results between the laser lancing device group and the conventional needle lancet group. The laser lancing device demonstrated comparatively lower pain than the conventional needle lancet. Lasers Surg. Med. © 2020 Wiley Periodicals LLC

 

 

 

 

https://onlinelibrary.wiley.com/doi/full/10.1002/lsm.23298

 